LX Washer Disinfector is a heavy-duty, high volume, single chamber, cabinet style washer designed for thorough and efficient cleaning of laboratory glassware, utensils, and miscellaneous items used in research and healthcare facilities.
LYNX 6100LX Surgical Instrument Washer SIZE
Compartment usable space: 26” wide x 26" high x 29" deep
Overall Unit: 42" wide x 80" high x 36" deep Sliding Door(s)
improved operating efficiencies
larger usable chamber size
Washer Disinfector CONFIGURATION (select one): front load and unload pass through load and unload The The LYNX 610 LX Surgical Instrument Washer is provided with an unload door and an additional cycle status indicator on the unload end, allowing one-way product flow from the contaminated area to the clean area.
Washer Disinfector shall be provided with a combination of rotating jet spray and direct spindle spraying for all treatment solutions. Washer Disinfector spray system consists of a single rotating spray arm mounted above, and interchangeable spindle wash or a rotating jet spray arms mounted below for internal treatment of the racked items. The bottom header automatically couples to the re-circulated wash/rinse treatment system.
Washer Disinfector VIEWING WINDOW
Washer door(s) shall be provided with a full view, tempered, double glass viewing window.
Surgical Instrument Washer Disinfector STAINLESS STEEL TREATMENT COMPONENTS
All treatment components, including jet spray system, screens, pump, and re-circulating piping are constructed of 304 stainless steel. Surgical Instrument Washer Disinfector PERSONNEL SAFETY FEATURES Opening the load-end door (or unload-end if pass-through unit), automatically stops operation of the washer. The door(s) must be closed, alarm acknowledged, and the “Cycle Start” button depressed to continue operation. An Emergency Stop (ES) button is provided on the load-end control panel (and unload-end if pass-through unit). If the operations of the wash cycle are interrupted by depressing the (ES) Button, the operator must reset the button and push the “Cycle Start” button to continue operations from the beginning of the selected cycle.
Washer Disinfector MULTI CYCLE COLOR/GRAPHIC MICRO COMPUTER CONTROL SYSTEM
A solid-state microcomputer control system monitors and automatically controls all process operations and functions. A twelve-cycle treatment menu may be programmed and retained to permit operating personnel to accommodate a wide variety of loads and processing requirements. Individual cycle phase times and temperatures and other key process parameters are programmable. Each cycle may be locked by supervision to ensure process integrity. The Cycle programming may be controlled by a supervisory access code. Each cycle program may be reviewed on demand.
washer disinfector instrument washer rinse temperatures
The standard treatment cycle consists of any or all of the following phases: Pre-wash, Wash, Rinse, Wash II, Rinse and Final Rinse. All cycle phases may be selected or deselected by the user. All Wash and Rinses treatments are re-circulated under pump pressure except the optional non-re-circulated rinse. The cycle, once activated, is completely automatic.
Located in the sump, easily removable screens are provided to protect items and large debris from entering the re-circulation system.
Water from the last rinse cycle may be held and utilized as the pre-wash of the next cycle.
Washer Disinfector is provided with a stainless steel trim flange to enclose the opening between one end of the washer and the masonry wall opening. A stainless steel modular wall assembly is designed and manufactured to enclose an open wall area or to create a wall barrier. All materials and fasteners are of stainless steel construction. A stainless steel transfer cart is provided to connect to the load or unload end of washer for accessories transportation and ease of loading of the washer.
improved operating efficiencies
larger usable chamber size
compact Surgical Instrument Washer Disinfectors cuts costs.